Pluse Oxymeter
MP-110PLUS






 
Display 240 x 64 Mono LCD
 
 
Display Paramers O2 Saturation, Pulser rate, BP(SYS/DIA/MEAN)
TEMP, SpO2 Wave
Battery / Sound / Alarm status
Time and Date
Tabular Trend
 
 
Alarm
 PR            High limit        30~ 250 bpm
                Low limit     20 ~245 bpm

SpO2        High limit        52 ~ 100%
                 Low limit         51 ~  99%

NIBP         SYS high/low limit   40~300/30~290 mmHg
                 DIA high/low limit    20~290/20~280 mmHg
                 MEAN high/low limit   35~295/25~285 mmHg
Temp        High limit       20 ~ 50℃
                 Low limit     10 ~ 40℃

Alarm silence           30 sec.
 
 
Power
 Power input     DC15V/3A
External DC     12~ 15V / 3A
Power consumption   24W max.
 
 
Battery
 12V/2Ah Pb-Acid
Full charging time  10Hrs max
Operating time      3Hrs max
 
 
Communication
 RS232C
 
 
Dimension
 W235 x H80 x 170 mm
 
 
Weight
 2.4Kg
 
 
Environmental
 Operationg temperature  10 ~40℃(50~104℉)
Storage Temperature     -20 ~ 70℃(-4 ~ 158℉)
Rerative humidity
Operating        10~90%
Storage      0~95%
 
 
printer(Optional)
 External type
58mm thermal printer